Radiotherapy/Nuclear Medicine

Á¦¸ñ Pro-NM NEMA NU2





Phantom for evaluating performance of positron emission tomographs (PET). Recommended for use in the evaluation of reconstructed image quality in whole body PET imaging. For simulation of whole-body imaging especially using PET and
camera-based coincidence imaging techniques.
Can also be used for determination of the coincidence count rate characteristics in brain and cardiac imaging, evaluation
of the relationship between true coincidence count rate and radioactivity, determination of the address errors caused by
address pile up, evaluation of the count loss correction scheme.

It is designed in accordance with the recommendations by the International Electrotechnical Commission (IEC) and modified by the National Electrical manufacturers Association (NEMA).

Product features:
Complies with:

– International Standard: Radionuclide imaging devices – Characteristics and test conditions
– Part 1: Positron emission tomographs, International Electrotechnical Commission (IEC), 61675-1, Geneva, Switzerland, 1998.
–Performance Measurements of Scintillation Cameras, NEMA Standards Publication No. NU2, National Electrical Manufacturers Association (NEMA), Washington, D.C., 2001.CE certified manual provides detailed guidelines for carrying oueach test, results assessment and registration

Technical data (can be modified to customer specifications):
interior length of phantom: 180 mm
volume of empty cylinder: 9.7 laters
6 fillable spheres - inner diameter: 10 mm, 13 mm, 17 mm, 22 mm, 28 mm, and 37 mm - distance from sphere
plane to inside wall: 70 mm
cylindrical insert dimension: - outside diameter: 51 mm - length: 180 mm