Radiotherapy/Nuclear Medicine

Á¦¸ñ Pro-NM PETscatter






The PET scatter phantom is acceptance testing tool used to determine the imaging systems relative sensitivity to scatter radiation. It can be used to measure the effects of dead-time and the effects of random events generated at different levels of activity of the line source.
The test phantom is a right circular cylinder composed of polyethylene. For ease of handling it consists of 3 segments that are assembled together during testing. A hole is drilled parallel to the central axis of the cylinder, at a radial distance of 45mm.

Technical data (can be modified to customer specifications):
• cylinder
- outside diameter: 203 mm
- length: 700mm
- hole diameter: 6.4 mm
- hole offset from the central axis: 45 mm
• line source
- outside diameter: 4.8 mm
- length: 800 mm
- inside diameter: 3.2 mm
• carrying case

Product features:
• Complies with: – NEMA Standard (NU 2-2007)
- NEMA 2012 Standard
• CE certified
• manual provides detailed guidelines for carrying out each test, results assessment and registration